AN012 - How to reset battery alarm on C1-455-B.

  • Get the programming cable built as shown:

  • Replace the battery with the instrument switched on.
  • Install the program "Teraterm Pro" after the extraction and follow the installation instructions.
  • Open the “Teraterm Pro”.
  • When opening the following screen will appear. Set the desired COM port.

  • Clock on the “Setup” menu and then select “Serial port”. Setting the speed to 19200.
  • Put in boot the C1-455 instrument, by pressing the “RESET” system button and now the “BOOT” button.
  • If the procedure is correct led “BOOT” lights.
  • Open the “Control” menu and select “Macro”. From the explorer window, select the file "ver.ttl" after extraction.

  • At this point the following screen will appear. Select “Yes”.

  • Set the date and follow the directions.
  • At this point the “BATT” led turns off.