AN012 - How to reset battery alarm on C1-455-B.
-
Get the programming cable built as shown:
-
Replace the battery with the instrument switched on.
-
Install the program "Teraterm Pro" after the extraction and follow the installation instructions.
-
Open the “Teraterm Pro”.
-
When opening the following screen will appear. Set the desired COM port.
-
Clock on the “Setup” menu and then select “Serial port”. Setting the speed to 19200.
-
Put in boot the C1-455 instrument, by pressing the “RESET” system button and now the “BOOT” button.
-
If the procedure is correct led “BOOT” lights.
-
Open the “Control” menu and select “Macro”. From the explorer window, select the file "ver.ttl" after extraction.
-
At this point the following screen will appear. Select “Yes”.
-
Set the date and follow the directions.
-
At this point the “BATT” led turns off.